Technical specifications of the units
Supply voltage 220-240 V; 50-60 Hz or 100-127 V; 50-60 Hz
Selectable via dummy plug located inside the unit by Technical Service
Power input with no HF power output: approx. 16 VA
at max. power output: approx. 800 VAa
Class of protection I
Classified acc. to MDD II b
Leakage currents LF and HF in acc. with EN IEC 60601, Part 2-2
Type of equipment CF; defibrillator-proof
Nominal frequency 450 kHz
Pulse frequency 30 kHz
HF output power:
Type of current Power Crest factor Voltage
Cutting 1 max. 400 W at 300 Ω 1.6 at 300 Ω max. 2300 Vpp
Cutting 2 max. 300 W at 300 Ω 1.9 at 300 Ω max. 2500 Vpp
Endo-Mode max. 100 W at 200 Ω 1.6 at 200 Ω max. 2800 Vpp
Contact coagulation max. 250 W at 200 Ω 3.4 at 200 Ω max. 3200 Vpp
Spray coagulation max. 120 W at 300 Ω 5.6 at 300 Ω max. 6000 Vpp
Bipolar coagulation max. 100 W at 100 Ω 2.1 at 100 Ω max. 600 Vpp
Duty type intermittent INT 10 s/30 s, equivalent to a duty factor of 25%
Mains fuses 220-240 V: T 4 A (slow-blow.)
100-127 V: T 8 A (slow-blow.)
Signal level HF activation: 55 dB (A) (adjustable between 50 dB and 60 dB by Technical Service)
Alarm: 65 dB(A)
Weight 8.6 kg
Interference suppression Limits in conformity with EN 55011, Interference immunity in conformity with IEC 801
Dimensions 405 mm x 135 mm x 380 mm (W x H x D)
CE marking conform with 93/42/EEC