Martin ME 411 Specifications
Supply voltage 100 V / 115 V / 127 V / 230 V / 240 V, 50 Hz - 60 Hz
Adjustable by the Technical Service
Power input in standby mode: 1 VA
with no HF power output: 42 VA
at max. power output: 880 VA
Class of protection I
Classified acc. to MDD IIb
Leakage currents in acc. with IEC 601, Part 2-2
Type of equipment CF; defibrillator-proof
Nominal frequency 4 00 kHz
Pulse frequencies 30 kHz at cut 2, URO cut 2 and contact coagulation 1
65 kHz at contact coagulation 2
50 kHz at spray coagulation
30 kHz at bipolar cutting 2
HF Power CF
Cutting 1 max. 320 W at 350 ohms 1.8 max. 2700 Vss
Cutting 2a max. 320 W at 350 ohms 2.3 max. 3500 Vss
Cutting 2b max. 320 W at 800 ohms 2.6 max. 4300 Vss
Urolog. cutting 1 max. 320 W at 350 ohms 1.8 max. 2700 Vss
Urolog. cutting 2 max. 320 W at 800 ohms 2.6 max. 4800 Vss
Contact coagulation 1a max. 250 W at 200 ohms 1.8 max. 1300 Vss
Contact coagulation 1b max. 250 W at 200 ohms 3.0 max. 2600 Vss
Contact coagulation 2 max. 150 W at 500 ohms 5.4 max. 4300 Vss
Spray coagulation max. 120 W at1000 ohms 5.5 max. 6000 Vss
Bipol. cutting 1 max. 80 W at 500 ohms 1.8 max. 900 Vss
Bipol. cutting 2 max. 80 W at 500 ohms 2.1 max. 1200 Vss
Bipol. coagulation max. 80 W at 100 ohms 1.8 max. 400 Vss
Duty type intermittent (10s/30s = duty factor of 25%)
Sound level HF activation 55 dB(A), (adjustable 50 dB - 60 dB by Technical Service)
Alarm: 65 dB(A)
Weight 14.2 kg
Interference suppression limits in conformity with EN 55011
interference immunity in conformity with IEC 801
CE marking conform with 93/42/EEC
Dimensions 405 mm x 135 mm x 380 mm (WxHxD)